510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1526-2015Class II

VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC

LivaNova USA, Inc. / initiated 2015-03-06 / Terminated / root cause: manufacturing process

Cyberonics, Inc began a recall of VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC on . FDA classified it as Class II and gives the reason as "The pulse generators have a lower battery longevity than specified in their design requirement as a result of the devices being inadvertently left in a programmed ON state during manufacture". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1526-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC.
Product code
LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
Reason for recall
The pulse generators have a lower battery longevity than specified in their design requirement as a result of the devices being inadvertently left in a programmed ON state during manufacture.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 03/6/2015, the recalling firm sent a recall notification letter to their consignees. The letter included an effectiveness card which is to be signed by the consignee and returned to the recalling firm.
Quantity in commerce
13 units
Distribution
AR, TX, GA, CA, MI, IN, FL, AZ, NV
Product codes and lots
Serial Numbers: 37627, 38473, 38539, 41446, 42887, 43491, 44186, 44339, 44391, 44443, 44455, 44510, 50715

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1526-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1526-2015Class IITerminatedVoluntary: Firm initiated