510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1630-2015Class II

Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic...

CooperSurgical, Inc. / initiated 2015-03-20 / Terminated / root cause: design

CooperSurgical, Inc. began a recall of Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic surgery on . FDA classified it as Class II and gives the reason as "The pad printing around the suture entry holes on the 15mm Suture Guide included in the kit may not have been appropriately manufactured resulting in the pad printing flaking off around the entry holes". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1630-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CooperSurgical, Inc.
Product
Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic surgery.
Product code
NLM Laparoscope, General & Plastic Surgery, Reprocessed
Reason for recall
The pad printing around the suture entry holes on the 15mm Suture Guide included in the kit may not have been appropriately manufactured resulting in the pad printing flaking off around the entry holes.
Root cause tag
design
FDA root cause
Component design/selection
Action
Consignees were notified of the recall by letter sent via FedEx dated March 20, 2015. The letter instructs consignees to do the following: discontinue use of the affected product, complete the attached acknowledgement and receipt form to return the product and schedule a replacement.
Quantity in commerce
430 units
Distribution
US Nationwide distribution.
Product codes and lots
Lot 174422

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1630-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1630-2015Class IITerminatedVoluntary: Firm initiated