PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is...
Bard Access Systems began a recall of PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is intended for use in the administration of fluids... on . FDA classified it as Class II and gives the reason as "Bard Access Systems is conducting a field action due to the potential that an incorrect Maximum Flow rate label associated to the power injection IFU for the PowerLoc MAX Safety Infusion Set (SWIS) was attached to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1560-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is intended for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. The PowerLoc MAX SWIS is also indicated for power injection of contrast media into the central venous system only with an implanted port that is also indicated for power injection.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Bard Access Systems is conducting a field action due to the potential that an incorrect Maximum Flow rate label associated to the power injection IFU for the PowerLoc MAX Safety Infusion Set (SWIS) was attached to the outside of the unit pouch.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- The firm, Bard, sent an "URGENT: Notification-Potential Incorrect Infusion IFU Label for PowerLoc MAX Safety Winged Infusion Set (SWIS)" letter dated March 25, 2015 to its customers. The letter described the product, problem and actions taken. The firm has corrected the cause of the issue and all future lots will be manufactured with the correct maximum flow rate of 2mL/second. Customers may return for a replacement or keep items. Please direct any return requests and questions to our coordinator at 801-522-5640.
- Quantity in commerce
- 700 units
- Distribution
- US Distribution to states of: GA and MN.
- Product codes and lots
- Lot D430107
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1560-2015 | Class II | Terminated | Voluntary: Firm initiated |