Z-1560-2015Class II

PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is...

Bard Access Systems, Inc. / initiated 2015-03-13 / Terminated / root cause: labeling and UDI

Bard Access Systems began a recall of PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is intended for use in the administration of fluids... on . FDA classified it as Class II and gives the reason as "Bard Access Systems is conducting a field action due to the potential that an incorrect Maximum Flow rate label associated to the power injection IFU for the PowerLoc MAX Safety Infusion Set (SWIS) was attached to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1560-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Access Systems, Inc.
Product
PowerLoc MAX Safety Winged Infusion Set, Product Code 0132215 The PowerLoc MAX Safety Winged Infusion Set is intended for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. The PowerLoc MAX SWIS is also indicated for power injection of contrast media into the central venous system only with an implanted port that is also indicated for power injection.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Bard Access Systems is conducting a field action due to the potential that an incorrect Maximum Flow rate label associated to the power injection IFU for the PowerLoc MAX Safety Infusion Set (SWIS) was attached to the outside of the unit pouch.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
The firm, Bard, sent an "URGENT: Notification-Potential Incorrect Infusion IFU Label for PowerLoc MAX Safety Winged Infusion Set (SWIS)" letter dated March 25, 2015 to its customers. The letter described the product, problem and actions taken. The firm has corrected the cause of the issue and all future lots will be manufactured with the correct maximum flow rate of 2mL/second. Customers may return for a replacement or keep items. Please direct any return requests and questions to our coordinator at 801-522-5640.
Quantity in commerce
700 units
Distribution
US Distribution to states of: GA and MN.
Product codes and lots
Lot D430107

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1560-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1560-2015Class IITerminatedVoluntary: Firm initiated