Z-1462-2015Class I

TIGERPAW System II, Part number C-TP-1509 (9 connector) The TigerPaw System II is an implantable occlusion...

Laax, Inc. / initiated 2015-03-25 / Terminated

Laax, Inc. began a recall of TIGERPAW System II, Part number C-TP-1509 (9 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left... on . FDA classified it as Class I and gives the reason as "Increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or may not be caused by incomplete closure...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1462-2015
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Laax, Inc.
Product
TIGERPAW System II, Part number C-TP-1509 (9 connector) The TigerPaw System II is an implantable occlusion Fastener to be placed on the Ostium of the Left Atrial Appendage (LAA).
Product code
GDW Staple, Implantable
Reason for recall
Increase of field reports involving issues with the TIGERPAW System II resulting in possible tissue tearing on left atrial wall and bleeding during use of the device, which may or may not be caused by incomplete closure of the TigerPaw System II fastener.
FDA root cause
Unknown/Undetermined by firm
Action
The firm, LAAx, Inc.on behalf of Maquet Medical Systems, sent an "URGENT MEDICAL DEVICE RECALL (Removal) IMMEDIATE ACTION REQUIRED " letter dated March 30, 2015 to all affected sites by express mail. The letter describers the product, problem and actions to be taken. The customers were instructed to examine your inventory immediately to determine if you have any TIGERPAW System II products; remove the products, quarantine them and place in a secure location; complete and return the MEDICAL DEVICE RECALL (Removal) RESPONSE FORM by e-mailing a scanned copy to:[email protected] or FAX to: 1-(973)-396-3607. If you have acted as a distributor for the TIGERPAW System product II, immediately contact those accounts, advise them of the recall situationa and have them return all stock to you. Your Marquet Medical System USA Sales Representative will contact you to assist with the return of any product. If you have any additional questions, contact your local Maquet Medical Systems USA Sales Representative or our Customer Service at 1-888-880-2874 (6:00 AM - 5:00 PM PST).
Quantity in commerce
~4,154 units distributed in US - both part numbers
Distribution
Worldwide Distribution: US (nationwide) and countries of:Australia, Bahrain, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy Netherland, Norway, Poland, Qatar, Slovenia, Spain, Saudi Arabia, Sweden, Switzerland, Turkey and United Kingdom.
Product codes and lots
Part number C-TP-1509 (9 connector)  All serial numbers/lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1462-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1461-2015Class ITerminatedVoluntary: Firm initiated
Z-1462-2015Class ITerminatedVoluntary: Firm initiated