EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used with an...
Olympus America Inc. began a recall of EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used with an Olympus video system center, light source... on . FDA classified it as Class II and gives the reason as "New and validated reprocessing procedures have been issued for the Olympus TJF-Q180V duodenoscope ("TJF-Q180V"), consisting of revised manual cleaning and high level disinfection procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2807-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- EVIS EXERA II Duodenovideoscope OLYMPUS TJF TypeQ180V. This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitors, EndoTherapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.
- Product code
- FDT Duodenoscope And Accessories, Flexible/Rigid
- Reason for recall
- New and validated reprocessing procedures have been issued for the Olympus TJF-Q180V duodenoscope ("TJF-Q180V"), consisting of revised manual cleaning and high level disinfection procedures.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- The firm, Olympus, sent an "Urgent Safety Notification" dated March 26, 2015, to direct consignees/customers. The notification described the product, problem and actions to be taken. The notification contained new reprocessing procedures. The customers were instructed to immediately implement new precleaning and high level disinfection procedure; ensure that all personnel involved in the reprocessing are completely knowledgeable and thoroughly trained on the new reprocessing instructions; complete and return the enclosed questionnaire via fax to (484) 896-7128 and implement the new cleaning instructions when the new brushes arrive on/about May 8, 2015. If you have question, contact our Technical Assistance Center (TAC) at 1-800-848-9024, option 1, Monday-Friday between 7 AM EST - 8 PM EST. For additional information on this matter, contact V.P., Regulatory/Clinical Affairs & Quality Assurance at (484) 896-5688 or by email at [email protected].
- Quantity in commerce
- 6109
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Canada and Mexico.
- Product codes and lots
- Model: TJF-Q180V, all sold units
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2807-2015 | Class II | Terminated | Voluntary: Firm initiated |