Z-1522-2015Class II

MicroScan -Neg Breakpoint Combo Panel Type 34; Catalog number B1017-404; Panel MicroScan Neg Breakpoint 34...

Siemens Healthcare Diagnostics Inc. / initiated 2014-05-14 / Terminated / root cause: manufacturing process

Beckman Coulter, Inc. began a recall of MicroScan -Neg Breakpoint Combo Panel Type 34; Catalog number B1017-404; Panel MicroScan Neg Breakpoint 34 Neg panels are designed for use in determining... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has received complaints regarding panel degradation with a portion of the MicroScan Neg Breakpoint Combo 34 (NBP34) panel lot ". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1522-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan -Neg Breakpoint Combo Panel Type 34; Catalog number B1017-404; Panel MicroScan Neg Breakpoint 34 Neg panels are designed for use in determining quantitative and /or qualitative antimicrobial agent susceptibility and /or identification to No.1 the species level of colonies, grown 34 on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive bacteria
Product code
JWY Manual Antimicrobial Susceptibility Test Systems
Reason for recall
Siemens Healthcare Diagnostics has received complaints regarding panel degradation with a portion of the MicroScan Neg Breakpoint Combo 34 (NBP34) panel lot 2014-05-28.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Siemens sent an Urgent Medical Device Recall letter dated May 14, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The risk to health is explained and users are asked to discontinue use and discard any of the recalled lot still in inventory. No other lot numbers of the NBPC 34 panels are affected and may still be used. The Field Correction Effectiveness Check form should have been completed and returned to Siemens. PLEASE FAX THIS COMPLETED FORM TO 302-631-8467. Contact Siemens Customer Care Center or local Siemens Tech support. For further questions please call (916) 374-3031
Quantity in commerce
681 boxes
Distribution
Worldwide Distribution - Nationwide Distribution including Puerto Rico and to the states of: KS, LA, MD, NY, WV,NM, NH, MO, CT, MA, OK, NC. IA, OH. WI, TX, PA, MS and MN and to the countries of : Canada, Kuwait, South Africa and Turkey.
Product codes and lots
Catalog Number B1017-404, lot # 2014-05-28, Expiry 2014-05-28.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1522-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1522-2015Class IITerminatedVoluntary: Firm initiated