510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1612-2015Class II

Fujifilm Synapse Cardiovascular I ProSolv CardioVascular Synapse Cardiovascular v4.0.8 Synapse Cardiovascular...

FUJIFILM Healthcare Americas Corporation / initiated 2015-02-04 / Terminated / root cause: software

Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm Synapse Cardiovascular I ProSolv CardioVascular Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1 Intended Use: Picture Archiving and... on . FDA classified it as Class II and gives the reason as "Under certain circumstances the patient Demographics in a report exported into the EMR may not match the demographics shown in the corresponding Synapse CV clinical report". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1612-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Fujifilm Synapse Cardiovascular I ProSolv CardioVascular Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1 Intended Use: Picture Archiving and Communication System Computer, Diagnostic, Programmable
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Under certain circumstances the patient Demographics in a report exported into the EMR may not match the demographics shown in the corresponding Synapse CV clinical report.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Fujifilm Medical on 2/4/14 , contacted users and advised to discontinue using HL7 reports pending FMSU determination of root cause. Firm issued follow-up letters on 2/24/14 and 7/24/14 advising users of a mandatory upgrade to version 5.1.1 I base, in-place upgrade, to correct the problems.
Quantity in commerce
16 units
Distribution
Nationwide Distribution including PR and the states of IL, IN, FL, GA, MD, MI, MO, MS, NC, NY, OR, and PA, PR and the country of Canada.
Product codes and lots
Software Versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1612-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1612-2015Class IITerminatedVoluntary: Firm initiated