Fujifilm Synapse Cardiovascular I ProSolv CardioVascular Synapse Cardiovascular v4.0.8 Synapse Cardiovascular...
Fujifilm Medical Systems U.S.A., Inc. began a recall of Fujifilm Synapse Cardiovascular I ProSolv CardioVascular Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1 Intended Use: Picture Archiving and... on . FDA classified it as Class II and gives the reason as "Under certain circumstances the patient Demographics in a report exported into the EMR may not match the demographics shown in the corresponding Synapse CV clinical report". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1612-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- Fujifilm Synapse Cardiovascular I ProSolv CardioVascular Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1 Intended Use: Picture Archiving and Communication System Computer, Diagnostic, Programmable
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Under certain circumstances the patient Demographics in a report exported into the EMR may not match the demographics shown in the corresponding Synapse CV clinical report.
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- Fujifilm Medical on 2/4/14 , contacted users and advised to discontinue using HL7 reports pending FMSU determination of root cause. Firm issued follow-up letters on 2/24/14 and 7/24/14 advising users of a mandatory upgrade to version 5.1.1 I base, in-place upgrade, to correct the problems.
- Quantity in commerce
- 16 units
- Distribution
- Nationwide Distribution including PR and the states of IL, IN, FL, GA, MD, MI, MO, MS, NC, NY, OR, and PA, PR and the country of Canada.
- Product codes and lots
- Software Versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1612-2015 | Class II | Terminated | Voluntary: Firm initiated |