FujiMedical Synapse Cardiovascular I ProSolv CardioVascular. A picture Archiving and Communication System...
Fujifilm Medical Systems U.S.A., Inc. began a recall of FujiMedical Synapse Cardiovascular I ProSolv CardioVascular. A picture Archiving and Communication System Software versions: Synapse Cardiovascular v4.0.8... on . FDA classified it as Class II and gives the reason as "Measurement values in the EMR may not accurately reflect the actual measurement value on the Synapse Cardiovascular clinical report". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1611-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Product
- FujiMedical Synapse Cardiovascular I ProSolv CardioVascular. A picture Archiving and Communication System Software versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Measurement values in the EMR may not accurately reflect the actual measurement value on the Synapse Cardiovascular clinical report.
- Root cause tag
- software
- FDA root cause
- Software in the Use Environment
- Action
- FujiMedical Systems North America issued intial recall letter dated February 4, 2014, advising users of the HL7 report exported to the EMR wherein measurement values presented in the Hl7 reports in your EMR may not accurately reflect the actual measurement value on the Synapse Cardiovascular clinical report. Follow-up letters were issued February 12, 2014, and July 24, 2014. A mandatory upgrade to software version 5.1.1/ base, in-place upgrade required. A customer response form attached to acknowledge receipt and arrange for the on site upgrade. For questions regarding this recall call 203-324-2000.
- Quantity in commerce
- 30 units
- Distribution
- Worldwide Distribution - US including CA, FL, IL, IN, MD, MI, MN, MO, NC, NY, OH, OR, PA, PR, TX, VA, WI Foreign: Canada
- Product codes and lots
- Software versions: Synapse Cardiovascular v4.0.8 Synapse Cardiovascular v4.0.8 SR1
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1611-2015 | Class II | Terminated | Voluntary: Firm initiated |