Z-1548-2015Class II

Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated...

Agilent Technologies, Inc. / initiated 2015-04-01 / Terminated / root cause: labeling and UDI

Dako North America Inc. began a recall of Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The... on . FDA classified it as Class II and gives the reason as "If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1548-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Agilent Technologies, Inc.
Product
Test Request Distributor (TRD 1.3 and TRD 1.4), a software module used on the Dako Omnis System, an automated slide stainer for in vitro diagnostic use. The TRD is intended to distribute patient case, appropriate parts information, track changes, and transform test requests from LIS or manual entries to connected Dako systems.
Product code
KPA Slide Stainer, Automated
Reason for recall
If a user requests slides from the LIS or TPID, then updates a request by changing the test, the TRD software will reject the update. The LIS and printed slide label will display the updated test with the requested change, but the Omnis system will execute the initial test, not reflecting the change. The system will not warn the user that the test request was rejected.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Dako sent a Voluntary Product Recall Notification letter starting the week of April 1, 2015 along with a acknowledgement attachment. The letter identified the affected product, problem and actions to be taken. The firm requests that customers inform those in their organization of the notification including organizations where the potentially affected instruments have been transferred. The firm requests that customers complete, sign, and return the acknowledgement form within one week after receiving the notification. For questions contact your Dako sales representative. On 04/29/15 the firm sent out an updated customer notification letter in order to provide additional clarification and instructions to customers in regards to the recall situation.
Quantity in commerce
33 (7 in U.S. and 26 Internationally)
Distribution
Worldwide Distribution - U.S Nationwide in the states of IN and PA and the countries of: Spain, Belgium, France, Netherlands, Denmark, Finland, Sweden, Germany, UK, and Korea.
Product codes and lots
N/A  Serial Number: 1311004352, 1308004862, 1307000537, 1405002266, 1410008844, 1307006995, 1311001075, 1402002186, 1401010308, 1402005995, 1306001697, 1311007317, 1410004066, 1404006200, 1410004067, 1405002263, 1404007225, 1410002144, 1311007321, 1404004226, 1401005960, 1305006010, 1401000919, 1405000371, 1401005963, 1403005465, 1304008205, 1401003711, 1401001048, 1310010425, 1401003715, 1311004354, 1311004355

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1548-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1548-2015Class IITerminatedVoluntary: Firm initiated