Z-1524-2015Class II

MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels...

Siemens Healthcare Diagnostics Inc. / initiated 2014-04-08 / Terminated / root cause: manufacturing process

Beckman Coulter, Inc. began a recall of MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed for use in determining quantitative and... on . FDA classified it as Class II and gives the reason as "Siemens investigation confirmed that a portion one lot of panel Pos Combo 33 lot was packaged without desiccants. The substrates on a panel packaged without a desiccant may deteriorate over time due to exposure to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1524-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
MicroScan Pos Combo Panel Type 33 (PC33): Catalog number B1017-211 MicroScan Pos Combo 33 B1017-211 panels are designed for use in determining quantitative and /or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive bacteria.
Product code
JWY Manual Antimicrobial Susceptibility Test Systems
Reason for recall
Siemens investigation confirmed that a portion one lot of panel Pos Combo 33 lot was packaged without desiccants. The substrates on a panel packaged without a desiccant may deteriorate over time due to exposure to additional moisture in the panel pouch.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens sent an Urgent Medical Device Recall letter dated April 8, 2014, to all affected customers. Letters advised users of the recalled product with catalog number, Siemens Material Number and Lot number as well as the reason for recall. The risk to health was explained as well as the recommendation for discussing the content of the letter with the medical director regarding the need to review previous test results. Customers are asked to discontinue use and discard any remaining inventory of the PC33 panel 2014-09-12. All other lot numbers are not impacted. The attached confirmation receipt is to be completed and returned and the letter should remain with laboratory records. Questions should be directed to Siemens Customer Care Center. For further questions call (916) 374-3031.
Quantity in commerce
664 panel boxes
Distribution
Nationwide Distribution including Puerto Rico and to the states of : CT, GA, NC, OK, NY, SC, KY, TN, CA, WA, OR, NJ, OH, PA, FL, MI, TX, WI, MO and NH.
Product codes and lots
Catalog number B10170211; Siemens Material number (SMN): 1071530, lot number 2014-09-12.: Expiration date 2014-09-12.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1524-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1524-2015Class IITerminatedVoluntary: Firm initiated