MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK began a recall of MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1) on . FDA classified it as Class II and gives the reason as "The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1574-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mitsubishi Heavy Industries, Ltd., Hiroshima Machinery Work
- Product
- MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- MITSUBISHI HEAVY INDUSTRIES, LTD. has issued a Safety Alert for the MHI-TM2000 Linear Accelerator System dated April 10, 2015, due to a software issue. Customers were advised not to deliver therapeutic X-ray if popup warning message show is displayed during treatment. Instructions were provided if therapeutic radiation is delivered during popup warning. Customers with questions should contact the Customer Service Representative by calling 1-800-597-5911, Monday through Friday or email: [email protected]. **CAPA** Correction software for this issue has a tentative availability date of June 2015. Estimated software update completion date would be end of October 2015.
- Quantity in commerce
- 11 units total
- Distribution
- Worldwide Distribution - US including FL, NY, OH, & TX, **Internationally - Belgium, Italy, Korea, France & Germany.
- Product codes and lots
- Serial Numbers: 201902, 201903, 202903, 202905, 203901, 203902, 203903, 203905, 203906, 203919 & 203924
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1574-2015 | Class II | Terminated | Voluntary: Firm initiated |