Philips Healthcare DuraDiagnost X- Ray
Philips Medical Systems, Inc. began a recall of Philips Healthcare DuraDiagnost X- Ray on . FDA classified it as Class II and gives the reason as "The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1555-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Healthcare DuraDiagnost X- Ray
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal. This is a failure to comply with 21CFR 1020.31(j).
- Root cause tag
- software
- FDA root cause
- Software design (manufacturing process)
- Action
- Philips Healthcare Planned action: 1. Philips healthcare will contact customers and initiate a software update to correct the software deficiency causing the problem. 2. Philips engineers will install the software. 3. Philips engineers will perform testing to ensure the software update was effective. 4. Corrections are to be made free of charge 5. Corrective actions should be completed by August 15, 2015 For further questions call (978) 659-7892
- Quantity in commerce
- 3 DuraDiagnost
- Distribution
- Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.
- Product codes and lots
- DuraDiagnost Stationary X-ray System.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1554-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1555-2015 | Class II | Terminated | Voluntary: Firm initiated |