Z-1743-2015Class II

VITROS 5600 Chemistry System, Software Version 3.2 & Below. Automates pre-analytical and post-analytical...

Ortho-Clinical Diagnostics, Inc. / initiated 2015-04-06 / Terminated / root cause: software

Ortho-Clinical Diagnostics began a recall of VITROS 5600 Chemistry System, Software Version 3.2 & Below. Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use on . FDA classified it as Class II and gives the reason as "Software anomaly allows testing of multiple assays using a single urine specimen, regardless of the pretreatment requirements for each selected assay, potentially leading to biased patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1743-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS 5600 Chemistry System, Software Version 3.2 & Below. Automates pre-analytical and post-analytical sample and data management for in vitro diagnostic use.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Software anomaly allows testing of multiple assays using a single urine specimen, regardless of the pretreatment requirements for each selected assay, potentially leading to biased patient results.
Root cause tag
software
FDA root cause
Software design
Action
On April 6, 2015, Ortho Clinical Diagnostics distributed correction notification letters and response forms to their customers via courier service. All urine patient results that require pretreatment was advised to refer to the notice of correction for instructions. Correction notices should also be posted beside the device user documentation for easy reference. Customers were advised to complete and return the Confirmation of Receipt before April 20, 2015. Customers with questions may contact the Customer Technical Service at 1-800-421-3311 anytime..***CAPA***Since the Assay Data Disk (ADD) provides data tables that determine which assays can be tested from the same sample container, an updated ADD is being created to address the anomaly on the VITROS 4600 and 5600. The anomaly will be addressed with an updated ADD is uploaded and the revised ADD mitigates the issue totally. Anticipated date of availability TBD.
Quantity in commerce
1895 units total (907 domestically & 988 internationally)
Distribution
Worldwide Distribution-US (nationwide), Brazil, Chile, Mexico, Australia, Hong Kong, India, Taiwan, Belgium, Denmark, France, Ireland, Italy, Portugal, Romania, Spain, Sweden and the UK.
Product codes and lots
Product Code: 6802413 Unique Device Identifier No.: 10758750002740

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1743-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1742-2015Class IITerminatedVoluntary: Firm initiated
Z-1743-2015Class IITerminatedVoluntary: Firm initiated