Z-1405-2015Class II

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx

GE Healthcare / initiated 2014-12-19 / Terminated

GE Healthcare began a recall of GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx on . FDA classified it as Class II and gives the reason as "An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1405-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx
Product code
IZL System, X-Ray, Mobile
Reason for recall
An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
FDA root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare planned action: 1. The customer notification letter will be distributed to the facilities that have installed the following AMX series X-ray systems: Models Brivo XR285 AMX, Optima XR200 AMX, and Optima XR220 AMX. 2. As stated in the customer notification letter, the correction will be provided free of charge and subsequent communication between the firm and the facility will result in a site visit by a GE Healthcare field engineer to bring the X-ray system into compliance. 3. The GEHC field engineer will follow Field Modification Instruction 10899 which provides instructions on how to perform a mAs accuracy check and how to determine if the test results pass predetermined acceptance criteria. If you have any questions or concerns regarding this notification, please call the Information following phone number: 1-800-437-1171.
Quantity in commerce
6 units US
Distribution
US Distribution to the states of: AZ, FL, GA, MO and WI.
Product codes and lots
BrivoXR285amx - PN 5555000, 5555000-2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1405-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1405-2015Class IITerminatedVoluntary: Firm initiated
Z-1406-2015Class IITerminatedVoluntary: Firm initiated
Z-1407-2015Class IITerminatedVoluntary: Firm initiated