Z-1407-2015Class II
GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
GE Healthcare / initiated 2014-12-19 / Terminated
GE Healthcare began a recall of GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx on . FDA classified it as Class II and gives the reason as "An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1407-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Healthcare planned action: 1. The customer notification letter will be distributed to the facilities that have installed the following AMX series X-ray systems: Models Brivo XR285 AMX, Optima XR200 AMX, and Optima XR220 AMX. 2. As stated in the customer notification letter, the correction will be provided free of charge and subsequent communication between the firm and the facility will result in a site visit by a GE Healthcare field engineer to bring the X-ray system into compliance. 3. The GEHC field engineer will follow Field Modification Instruction 10899 which provides instructions on how to perform a mAs accuracy check and how to determine if the test results pass predetermined acceptance criteria. If you have any questions or concerns regarding this notification, please call the Information following phone number: 1-800-437-1171.
- Quantity in commerce
- 6 units in US
- Distribution
- US Distribution to the states of: AZ, FL, GA, MO and WI.
- Product codes and lots
- OptimaXR220amx PN 5555000-5, PN5555000-6
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1405-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1406-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1407-2015 | Class II | Terminated | Voluntary: Firm initiated |