VasoNova Vascular Positioning System (VPS ) G4 Console Catalog number VPS-G4C Cardiovascular: The VPS Stylet...
VasoNova, Inc. began a recall of VasoNova Vascular Positioning System (VPS ) G4 Console Catalog number VPS-G4C Cardiovascular: The VPS Stylet and Console are indicated for guidance and tip... on . FDA classified it as Class II and gives the reason as "A limited number of VPS G4 Consoles are being recalled due to noncompliance with International Electrotechnical Commission Standard 60101-1-1". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1408-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- VasoNova Vascular Positioning System (VPS ) G4 Console Catalog number VPS-G4C Cardiovascular: The VPS Stylet and Console are indicated for guidance and tip positioning for central venous catheters.
- Product code
- OBJ Catheter, Ultrasound, Intravascular
- Reason for recall
- A limited number of VPS G4 Consoles are being recalled due to noncompliance with International Electrotechnical Commission Standard 60101-1-1.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Teleflex sent an Urgent Medical Device Recall Notification letters dated on March 30, 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. The risk to users is identified and users are asked to discontinue use and quarantine affected serial numbers. Teleflex sales reps will help with return of product. The Acknowledgeable form should be completed and returned as per instructions. Questions should be directed to 1.877.236.6869.
- Quantity in commerce
- 17
- Distribution
- US Nationwide Distribution in the states of AZ, DC, PA, IN, PA. GA, NY, KS, CA, VA, NC.
- Product codes and lots
- Product code VPS-G4C; Serial numbers: GB40129, GB40130, GB40131, GB40133, GB40136, GB40146, GB40148, GB40151, GB40152, GB40154, GB40155, GB401456, GB40157, GB40158, GB40168, GB40175, GB40186
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1408-2015 | Class II | Terminated | Voluntary: Firm initiated |