Z-1561-2015Class II

AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation. AMSCO C Small Sterilizer, AMSCO 400 Small...

STERIS Corporation / initiated 2015-02-19 / Terminated / root cause: software

Steris Corporation began a recall of AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation. AMSCO C Small Sterilizer, AMSCO 400 Small Sterilizer, AMSCO 400 Medium Sterilizer. Designed for... on . FDA classified it as Class II and gives the reason as "STERIS has identified that the control board software in select AMSCO 400 and AMSCO C units will interrupt and cancel a processing cycle should the selected sterilizer shutdown time coincide with a processing cycle...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1561-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation. AMSCO C Small Sterilizer, AMSCO 400 Small Sterilizer, AMSCO 400 Medium Sterilizer. Designed for sterilization of heat and moisture-stable materials used in healthcare facilities.
Product code
FLE Sterilizer, Steam
Reason for recall
STERIS has identified that the control board software in select AMSCO 400 and AMSCO C units will interrupt and cancel a processing cycle should the selected sterilizer shutdown time coincide with a processing cycle. This may result in a procedure delay.
Root cause tag
software
FDA root cause
Software design (manufacturing process)
Action
On 3/9/2015, the firm sent Urgent Field Correction Notices, to their customers. The letter identified the affected product and gave a description of the problem. The letter states that the control software is being revised to ensure the utility shutdown feature will not interrupt an ongoing processing cycle. STERIS Field Representatives will contact customers to schedule installation of the new control board on affected units. Customers may direct questions regarding the issue to STERIS Customer Service at 1-800-548-4873; Marie LaFrance, Product Manager, at 440-392-7648, or their local sales representative.
Quantity in commerce
701 units
Distribution
Worldwide Distribution -- US, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, I, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY; and, the countries of Afghanistan, Canada, Japan, Kuwait, Mexico, Republic of Korea, & United Arab Emirates.
Product codes and lots
AMSCO 400 Units: Serial #'s: 030051501 - 033641440; AMSCO C Units: Serial #'s: R020051501 - R023641403.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1561-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1561-2015Class IITerminatedVoluntary: Firm initiated