Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1597-2015Class III

Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a...

Spinal Elements, Inc / initiated 2015-04-09 / Terminated

Spinal Elements, Inc began a recall of Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a variety of screws, hooks, rods, connectors, and... on . FDA classified it as Class III and gives the reason as "Some rods may not meet the appropriate traceability requirements including the lot number. Some rods were found to not include the product identification number (product number and lot number) marked directly on the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1597-2015
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Spinal Elements, Inc
Product
Spinal Elements 5.5 x 45MM Mercury lordosed rod. A component of the Mercury Spinal System, comprised of a variety of screws, hooks, rods, connectors, and staples that are used for attachment to the non-cervical spine. Rods span the distance between screws and hooks and achieve fixation by the mechanical joining of the rods with the screws or hooks. Lordosed rods are pre-curved to better fit the patient anatomy.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Some rods may not meet the appropriate traceability requirements including the lot number. Some rods were found to not include the product identification number (product number and lot number) marked directly on the rod. The lack of marking interrupts the ability to maintain traceability of the device.
FDA root cause
Under Investigation by firm
Action
The firm sent customer notification letters on 04/09/15. In this letter the firm requested that all affected lots be returned, and replacement inventory will be made available. The firm requested that customers take the following actions: 1. Immediately examine inventory for part number 60100-045. 2. Immediately discontinue the use and/or distribution of the affected lots 3. Send all affected product back to: Spinal Elements, Inc. ATTN: Julie Lamothe 3115 Melrose Drive Carlsbad, CA 92010 4. If further distributed the product to please identify the customers and notify them as once. 5. Complete and return the enclosed "Customer Response Form." Any questions the firm states to contact the Customer Service department at 1.877.SPINAL5 Ext. 213.
Quantity in commerce
588
Distribution
Nationwide Distribution including the states of GA, CA, TX, CO, IN, NC, FL, UT, TN, MO, MI, PA, OH, and OK.
Product codes and lots
Part Number: 60100-045 Lot Numbers: 140086, 140449, 141013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1597-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1597-2015Class IIITerminatedVoluntary: Firm initiated