Z-1675-2015Class II

Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm...

Datascope Corp. / initiated 2015-03-12 / Terminated / root cause: software

Mindray DS USA, Inc. dba Mindray North America began a recall of Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network... on . FDA classified it as Class II and gives the reason as "Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1675-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Panorama Patient Monitoring Network, Multi- Parameter Patient Monitor (with Arrhythmia Detection and Alarm. Part number 0998-00-0708-01. The Panorama Network includes the Panorama Telemetry System, which acquires and monitors physiological data for ambulating patients within a defined coverage area.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Software anomaly in the Panorama System software version 8.9 that manifests when the Panorama is in use with Mindray Passport V Monitor. Arrhythmia alarms which were previously displayed on both the Passport V and Panorama will not be displayed on the Panorama following a restart of either system (due to communication loss, restart, etc.).
Root cause tag
software
FDA root cause
Software design
Action
Mindray sent a Corrective Action letter dated March 12, 2015 to their affected customer via FedEx. The letter identified the affected product, problem and actions to be taken. For questions call 201-995-8407.
Quantity in commerce
4 units
Distribution
Worldwide Distribution - US Nationwide in the state of TX.
Product codes and lots
Part No: 0998-00-0708-01 Software version 8.9

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1675-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1675-2015Class IITerminatedVoluntary: Firm initiated