Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only...
Biomet Spine, LLC began a recall of Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing... on . FDA classified it as Class II and gives the reason as "Multiaxial screws may not meet internal requirements related to fatigue strength. Compression testing result and surface treatment are out of specification". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2152-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Highridge Medical, LLC
- Product
- Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing Technologies. Spinal fixation component.
- Product code
- NKB Thoracolumbosacral Pedicle Screw System
- Reason for recall
- Multiaxial screws may not meet internal requirements related to fatigue strength. Compression testing result and surface treatment are out of specification.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Biomet Spine issued an Urgent Medical Device Recall Notice/Response Form dated 3/12/2015 to their affected customers.
- Quantity in commerce
- 4,765 units
- Distribution
- Distribution US Nationwide (AL, CA, CO, FL, GA, LA, MA, MD, MI, MO, NY, OH, PA, TX, WA, WI) and The Netherlands.
- Product codes and lots
- Part Numbers: 14-582020, 14-582022, 14-582025, 14-582030, 14-582035, 14-582040, 14-582120, 14-582122, 14-582125, 14-582127, 14-582130, 14-582135, 14-582140, 14-582220, 14-582222, 14-582225, 14-582230, 14-582235, 14-582240, 14-582245, 14-582250, 14-592320, 14-582325, 14-582330, 14-582335, 14-582340, 14-582345, 14-582350, 14-582530, 14-582535, 14-582540, 14-582545, 14-582550, 14-582565, 14-582570, 14-582575, 14-583020, 14-583025, 14-583030, 14-583035, 14-583040, 14-583120, 14-583125, 14-583130, 14-583135, 14-583140, 14-583220, 14-583225, 14-583230, 14-583235, 14-583240, 14-583245, 14-583250, 14-583320, 14-583325, 14-583330, 14-583335, 14-583340, 14-583345 and 14-583350. All lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2152-2015 | Class II | Terminated | Voluntary: Firm initiated |