Z-1661-2015Class II

Direct Water feed Kit, Catalog#10373222

Siemens Healthcare Diagnostics Inc. / initiated 2015-03-05 / Terminated / root cause: manufacturing process

Siemens Medical Solutions Diagnostics began a recall of Direct Water feed Kit, Catalog#10373222 on . FDA classified it as Class II and gives the reason as "The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1661-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Direct Water feed Kit, Catalog#10373222
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens Healthcare Diagnostics sent an Urgent Device Correction Letter/Field Correction Effective Check Form dated 3/5/15 via FedEx to their affected customers.
Quantity in commerce
15 bottles
Distribution
Worldwide Distribution.
Product codes and lots
Direct Water Feed Kit - Product Code/Catalog #10373222

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1661-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1658-2015Class IITerminatedVoluntary: Firm initiated
Z-1659-2015Class IITerminatedVoluntary: Firm initiated
Z-1660-2015Class IITerminatedVoluntary: Firm initiated
Z-1661-2015Class IITerminatedVoluntary: Firm initiated
Z-1662-2015Class IITerminatedVoluntary: Firm initiated
Z-1663-2015Class IITerminatedVoluntary: Firm initiated