Z-1935-2015Class II

Ocean Water Seal Chest Drain, Single Collection W/AC, P/N: 2002-000

Atrium Medical Corporation / initiated 2015-05-07 / Terminated / root cause: manufacturing process

Atrium Medical Corporation began a recall of Ocean Water Seal Chest Drain, Single Collection W/AC, P/N: 2002-000 on . FDA classified it as Class II and gives the reason as "Chest drains have a missing o-ring". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1935-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Atrium Medical Corporation
Product
Ocean Water Seal Chest Drain, Single Collection W/AC, P/N: 2002-000
Product code
PAD Chest Drainage Kit
Reason for recall
Chest drains have a missing o-ring
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Atrium Medical issued recall letter dated 5/7/15 to inform customers of the steps to take should they encounter affected product. If the O-ring is not present on the patient tubing: a. Do not use the chest drain. b. Replace the entire chest drain including the patient tubing. Open up another chest drain package and repeat steps 1 to 3. c. Return any units with missing O-rings to Atrium. If the O-ring is present then reconnect in-line connector and proceed with set up as normal. Complete and return Field Action Reply Form to the following e-mail address: [email protected] or you may fax it to 1-603-386-6590. If you have any questions, please contact your local Atrium/Maquet chest drainage representative or Atrium Customer Service at 1-800-528-7486 (603-880-1433), Monday through Friday between 8:00 am and 5:00 pm (EST).
Quantity in commerce
390 cases
Distribution
Nationwide Canada, Germany
Product codes and lots
Lot Number: 222511

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1935-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1935-2015Class IITerminatedVoluntary: Firm initiated
Z-1936-2015Class IITerminatedVoluntary: Firm initiated
Z-1937-2015Class IITerminatedVoluntary: Firm initiated
Z-1938-2015Class IITerminatedVoluntary: Firm initiated