COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in...

Siemens Healthcare Diagnostics Inc. / initiated 2015-04-13 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting... on . FDA classified it as Class II and gives the reason as "Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1616-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.
Product code
GIF Diluent, Blood Cell
Reason for recall
Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent Medical Device Recall letter dated 4/13/15 was sent to customers who purchased the Coulter DxH Diluent. The letter informs customers that Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The letter informs customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Center via http://www.beckmancoulter.com/customersupport/support or call (800) 526-7694 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
Quantity in commerce
829,888 units total (659,584 units in US)
Distribution
Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and Tobago; United States, including Puerto Rico; and Uruguay.
Product codes and lots
Lot No. 3099980 - 3512190, expiration dates: 17 April 2015 - 26 August 2016.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1616-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1616-2015Class IITerminatedVoluntary: Firm initiated