Z-1604-2015Class I
HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit
HeartWare, Inc. / initiated 2015-04-29 / Open, Classified / root cause: design
HeartWare Inc began a recall of HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been... on . FDA classified it as Class I and gives the reason as "Failures of the splice repair kit if exposed to excessive force". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1604-2015
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- HeartWare, Inc.
- Product
- HeartWare Ventricular Assist System (HVAD) Pump Driveline Splice Kit Product Usage: Is intended to repair the driveline cable once an electrical break has been identified. These repairs are limited to the portion of the driveline that is outside the patient boundary (exit side).
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- Failures of the splice repair kit if exposed to excessive force.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- HeartWare, Inc. sent an Urgent Medical Device Correction letter to all affected consignees on May 11, 2015. The letter identified the affected product, problem, and actions to be taken by the consignees. The firm requested that consignees: 1) review the notice and the Patient Communication contents, 2) forward the notice to individuals within their organization who need to be aware of the notice, 3) identify patients that are currently on support and have undergone a driveline splice repair, 4) distribute the Patient Communication to the affected patients in person, 5) continue to reinforce the messages described in the notice with patients who have experienced a splice repair during their regularly scheduled appointments, 6) complete, sign, and return the "Acknowledge and Completion Form to HeartWare within 30 days of receipt of the letter. Should any questions or concerns arise, please contact there local HeartWare representative. Their 24-hour Clinical Support personnel are also available at 1 (888) 494-6365.
- Quantity in commerce
- 35
- Distribution
- Worldwide Distribution - US Nationwide in the states of FL, KY, MO, NY, TN,TX, and the countries of: Canada, Germany, Italy, Netherlands, Turkey and United Kingdom.
- Product codes and lots
- Catalog #: ASY00116: HVAD Pump Driveline Splice Kit Lot #'s 532770, 567949, 480786, 483769, 490692, 539664, and 598189 and Catalog #: ASY00281: HVAD Pump Driveline Splicekit; Large OD Sheath Lot #'s 654090, 672401, 678800, 690423, 833963, 873159, and 880816