Z-1643-2015Class II

Pinnacle3 Software Version 10.0, Model 459800200841

Philips North America LLC / initiated 2015-04-23 / Terminated / root cause: software

Philips Medical Systems, Inc. began a recall of Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data... on . FDA classified it as Class II and gives the reason as "A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1643-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Pinnacle3 Software Version 10.0, Model 459800200841. Product Usage: The Pinnacle 3 RTP software allows qualified medical professionals to enter patient data into the system, use that data to construct a plan for radiation therapy and evaluate the plan. Optionally, the qualified medical personnel may output the plan in an electronic or printed form for use by other systems in the delivery of treatment to a patient.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
A problem (the dose computed in planning mode is calculated incorrectly ) has been detected in the Philips Pinnacle3 software version 10.0 that, if it were to re-occur, could pose a risk for patients or users. Specifically, the dose engine is being passed the wrong snout position. The snout position is used to determine the penumbra of the beam which includes the calculation of the source si
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare sent a "Field Safety Notification PROST-T-300-01-06" letter dated 20 April 2015 to their customers. The letter described the Problem Description, Hazard Involved, How to Identify Affected Products, Action to be Taken by Customer / User, Actions Planned by Philips and Further Information and Support (Contact Philips Customer Care Solutions Center at 1-800-722-9377).
Quantity in commerce
4
Distribution
US Nationwide Distribution in the states of MO, FL, OH
Product codes and lots
"System Code S/N"  1439NML05A, 1117FMM025, DC090775040, 1117FMM03J.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1643-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1643-2015Class IITerminatedVoluntary: Firm initiated