Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1605-2015Class II

Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only, Product Usage Usage: Surgical instruments used in...

BioHorizons Implant Systems Inc / initiated 2015-04-16 / Terminated / root cause: manufacturing process

BioHorizons Implant Systems Inc began a recall of Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only, Product Usage Usage: Surgical instruments used in preparation for dental implant placement on . FDA classified it as Class II and gives the reason as "BioHorizons Tapered HD Upgrade set was packaged with an incorrect drill component manufactured of stainless steel in place of the correct component manufactured from titanium". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1605-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BioHorizons Implant Systems Inc
Product
Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only, Product Usage Usage: Surgical instruments used in preparation for dental implant placement
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
BioHorizons Tapered HD Upgrade set was packaged with an incorrect drill component manufactured of stainless steel in place of the correct component manufactured from titanium.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
BioHorizons sent an Urgent Medical Device Field Correction letter on 04/16/2015 to affected customers. Non-responding customers will be notified after 30 days. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Care Department at 888-246-8338.
Quantity in commerce
48 units
Distribution
US Nationwide in the states of: AL, FL, MS, VA, LA, CA, NC, TX, TN, OH, OK, VT, WA, MD, NY, SD, and SC.
Product codes and lots
Lot Number 1501968

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1605-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1605-2015Class IITerminatedVoluntary: Firm initiated