Z-1599-2015Class II
VenaSeal, Sapheon Closure System, Model # SP-101, VS-301 and VS-402. For vascular use
Medtronic, Inc. / initiated 2015-04-28 / Terminated / root cause: packaging
Covidien LLC began a recall of VenaSeal, Sapheon Closure System, Model # SP-101, VS-301 and VS-402. For vascular use on . FDA classified it as Class II and gives the reason as "Possible sterility breach of the outer packaging (pouch) material which may compromise the outside surface sterility of the sealed inner tray". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1599-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- VenaSeal, Sapheon Closure System, Model # SP-101, VS-301 and VS-402. For vascular use.
- Product code
- PJQ Agent, Occluding, Vascular, Permanent
- Reason for recall
- Possible sterility breach of the outer packaging (pouch) material which may compromise the outside surface sterility of the sealed inner tray.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Consignees were contacted by letter via email on April 28, 2015.
- Quantity in commerce
- 1661 units
- Distribution
- Worldwide distribution. No U.S. accounts; Australia, Canada, France, Germany, Hong Kong, Ireland, Italy, Latvia, Netherlands, Norway, Portugal, Switzerland, UAE and the United Kingdom.
- Product codes and lots
- Model no. SP-101, Lots, 35766, 36191, 36192, 36617, 36618, 37621, 37622, 37399, 37400, 38263, 38286, 38407, 38417, 38756, 38863, 38864, 39113, 39336, 39378, 39670, 39739, 39848 and 40075.; Model no. VS-301, Lots 34193, 35585, 37007, 38804, 39111, 39123, 39469 and 40160; Model no. VS-402, Lots 39970, 39971, 39972, 39973, 40067, 40069, 40187, 40243, 40271 and 40272.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1599-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.