Z-1962-2015Class II

The Rapidlab 1265 system is a blood gas analyzer used for laboratory testing of blood gases, electrolytes...

Siemens Healthcare Diagnostics Inc. / initiated 2015-05-06 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc began a recall of The Rapidlab 1265 system is a blood gas analyzer used for laboratory testing of blood gases, electrolytes, metabolites and CO-oximetry in arterial, venous and... on . FDA classified it as Class II and gives the reason as "D50 and D51 diagnostic error codes are not functional". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1962-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
The Rapidlab 1265 system is a blood gas analyzer used for laboratory testing of blood gases, electrolytes, metabolites and CO-oximetry in arterial, venous and capillary whole blood samples.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
D50 and D51 diagnostic error codes are not functional.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent an Urgent Field Safety Notice April 2015, to all affected customers. Customers were notified of the recall beginning on May 6, 2015 via Federal Express. Foreign customers were notified starting on May 5, 2015 via method determined by each countries local regulations and procedures. The notification informs customers of the issue and asks customers to do the following: Ensure that the Glucose and Lactate sensor contacts are aligned and completely seated with the contacts on the analyzer, Analyze a minimum of 2 levels of quality control material after installing the sensors, as indicated in the RAPIDLab 1200 Operator's Guide, Review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 7 days. Retain this letter with your laboratory records, and forward this letter to those who may have received this product. For further questions please call (800) 242-3233.
Quantity in commerce
2713 devices
Distribution
Worldwide Distribution - Nationwide Distribution including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Bosnia, Herzog,Brazil, Brunei, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Finland, France, Germany,Georgia, Greece, Hong Kong, Iceland, India, Indonesia, Ireland, Italy, Japan, Latvia, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, P.R. China, Peru, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., United Kingdom, Venezuela and Vietnam.
Product codes and lots
Siemens Material Number (SMN): 10321852, 10470366, 10491395, 10335524

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1962-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1961-2015Class IITerminatedVoluntary: Firm initiated
Z-1962-2015Class IITerminatedVoluntary: Firm initiated