CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT...
GE Healthcare began a recall of CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state detector on . FDA classified it as Class II and gives the reason as "GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1623-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state detector.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 25461 dated April 22, 2015. The letter was addressed to Hospital Administrators / Risk Managers, Managers of Radiology/Cardiology, and Radiologists/Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. The letter stated that customers can continue to use their device. If the error occurs, then stop using the system and call a service representative to bring the system back into proper operation. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. UPDATED: Consignees were sent on 5/14/2015 a second GE Healthcare "Urgent Medical Device Correction" letter dated May 14th, 2015. The letter was to incorporate and additional unit being recalled.
- Quantity in commerce
- 4 USA
- Distribution
- Distributed to the states of MN, FL & TX.
- Product codes and lots
- Mfg Lot or Serial # System ID 00000331100CN8 904223CT 00000345975CN7 409838LS16 00000337343CN8 512OCCT 00000344962CN6 INSIGHTPCT1356
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1623-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
