510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1700-2015Class II

Swan-Ganz CCO/Sv02/CEDV/VIP Thermodilution Catheters model number 744HF75. Flow-directed pulmonary artery...

Edwards Lifesciences LLC / initiated 2015-05-05 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of Swan-Ganz CCO/Sv02/CEDV/VIP Thermodilution Catheters model number 744HF75. Flow-directed pulmonary artery catheters for monitoring cardiac output, mixed venous... on . FDA classified it as Class II and gives the reason as "The product failed to meet the internal established statistical test requirement for heparin activity". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1700-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
Swan-Ganz CCO/Sv02/CEDV/VIP Thermodilution Catheters model number 744HF75. Flow-directed pulmonary artery catheters for monitoring cardiac output, mixed venous oxygen saturation, and end diastolic volume when used with the Vigilance Monitor.
Product code
DYG Catheter, Flow Directed
Reason for recall
The product failed to meet the internal established statistical test requirement for heparin activity.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On 05/05/2015 Edwards sent their consignees a letter including the following instructions: Once you have verified your inventory, please complete the attached acknowledgement form (even if you have no inventory) and fax it to Edwards Customer Service at 1-800-422-9329 within three days of receipt of this notification. This is to confirm that you have reviewed this notice and have taken appropriate action. Please call Edwards Customer Service at 1-800-424-3278 to arrange for return of any unused product from the identified lot and obtain information about replacement product. Please call for an RGA prior to returning any unused product and return unused product to the following address: Edwards Lifesciences Return Goods/RGA # XXXXXX 1212 Alton Pkwy., Irvine, CA 92606-4838
Quantity in commerce
978
Distribution
Worldwide distribution US nationwide, Canada and Europe (IC, BE, IT, GB, FR, DK, CZ, DE, AT, ES, IE, NO, PL)
Product codes and lots
lot number 59860969.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1700-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1700-2015Class IITerminatedVoluntary: Firm initiated