Mucolytic Agent 1 pint/473 ml; 1 gallon/3.8 l
Richard-Allan Scientific Company began a recall of Mucolytic Agent 1 pint/473 ml; 1 gallon/3.8 l Product Usage: A mucoliquefying preservative designed for use in the preparation of mucoid cytology specimens... on . FDA classified it as Class II and gives the reason as "Richard Allan Scientific (RAS) is initiating a recall of Mucolytic agent and Mucolexx following a customer complaint reporting that a lab had found fungal contamination in 2 lots of Mucolexx". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1639-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Richard-Allan Scientific LLC
- Product
- Mucolytic Agent 1 pint/473 ml; 1 gallon/3.8 l Product Usage: A mucoliquefying preservative designed for use in the preparation of mucoid cytology specimens. Mucolexx & Mucolytic Agent liquefies mucus and allows for optimal cytologic diagnostic results.
- Product code
- LEA Preservative, Cytological
- Reason for recall
- Richard Allan Scientific (RAS) is initiating a recall of Mucolytic agent and Mucolexx following a customer complaint reporting that a lab had found fungal contamination in 2 lots of Mucolexx.
- Root cause tag
- manufacturing process
- FDA root cause
- Material/Component Contamination
- Action
- Richard Allan Scientific (RAS) initiated a voluntary recall of Mucolexx and Mucolytic Agent due to fungal contamination on 5/1/2015 via certified mail to all customers who received the affected. Customers were asked to return a repsonse form, and dispose or return all recalled products. Please contact Sarah Rickert at 269-544-5628 or sarah.rickert@thermofisher
- Quantity in commerce
- Product 76166- 116; Product Code 76167-29
- Distribution
- Worldwide Distribution - US Nationwide in the states of: AR; AZ; CA; CO; FL; GA; I A; IL; IN; KY; MA; MN; MO; MS; MT; NC; ND; NJ; NV; NY; OH; OK; PA; PR; SD; TN; TX; VA; WA; WI and the countries of: Bangladesh, Canada, China, Costa Rica, Dubai, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Korea, Netherlands, Panama, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, and UK.
- Product codes and lots
- Richard Allan Scientific Mucolytic Agent 1 pint / 473 ml Product code: 76166, Lot #s and EXP Dates: 313839-Dec 2016; 311157-Oct-2016; 306726-Aug-2016; 297142-Jun-2016; 296289-May-2016; 290486-Mar-2016; 289488-Mar-2016; 276824-Nov-2015; 271110-Oct-2015; 265368-Aug-2015. Richard Allan Scientific Mucolytic Agent 1 gallon/ 3.8 l, Product Code: 76167; Lot Number & EXP Dates: 311158- Oct-2016; 302099- Jun-2016; 298834-Jun-2016; 293734- Apr-2016; 288009-Feb-2016; 283396-Jan-2016; 279581-Nov-2015; and 274814- Oct-2015 Dec-2016 Oct-2016 Aug-2016 Jun-2016 May-2016 Mar-2016 Mar-2016 Nov-2015 Oct-2015 Aug-2015
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1639-2015 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1640-2015 | Class II | Terminated | Voluntary: Firm initiated |