510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1642-2015Class II

DeRoyal(R) Medium Specimen Retrieval Bag, REF 5500-000-003, Rx Only

DeRoyal Industries, Inc. / initiated 2015-04-17 / Terminated / root cause: design

DeRoyal Industries Inc began a recall of DeRoyal(R) Medium Specimen Retrieval Bag, REF 5500-000-003, Rx Only Product Usage: The device is a single-use specimen retrieval bag indicated for use in... on . FDA classified it as Class II and gives the reason as "The firm received reports of specimen retrieval bags tearing, ripping, and coming apart during use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1642-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DeRoyal Industries, Inc.
Product
DeRoyal(R) Medium Specimen Retrieval Bag, REF 5500-000-003, Rx Only Product Usage: The device is a single-use specimen retrieval bag indicated for use in laparoscopic procedures to capture and remove organs or tissue from the body cavity.
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
The firm received reports of specimen retrieval bags tearing, ripping, and coming apart during use.
Root cause tag
design
FDA root cause
Device Design
Action
DeRoyal sent an Urgent: Voluntary Recall letter dated April 17, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to 1. Check your inventory for the product and specific lot number from the attached Recall Response Form. 2. If you have affected inventory, please destroy it and dispose of it according to your facilitys procedures. 3. Record all information on the included Recall Response and Notice of Destruction forms. 4. If you purchase through a distributor, please indicate their name/location on the Notice of Destruction Form so that we may issue credit appropriately. 5. Return both forms to [email protected] or 865-362-3716 (fax) even if you no longer have affected inventory. 6. Once we have received your Recall Response and Notice of Destruction forms, you will receive credit for destroyed product as indicated on the form. 7. If you are a distributor of these products, please forward this recall notice on to your end-users, or provide us with your customer listing and we will gladly contact them for you. For questions regarding this recall, please contact your DeRoyal Sales Representative or DeRoyal Customer Service at 1-800-251-9864.
Quantity in commerce
5359 units
Distribution
US Nationwide Distribution
Product codes and lots
1211QDA403, 1301QDA402, 1301QDA410, 1303QDA401, 211QDA403, 32046564, 32119911, 32338381, 32562251, 32783812, 32886441, 32891051, 32899882, 32942609, 33114274, 34939294, 35229930, 35579999, 35738515

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1642-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1642-2015Class IITerminatedVoluntary: Firm initiated