Z-1828-2015Class II

Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit...

Terumo Cardiovascular Systems Corporation / initiated 2015-05-06 / Terminated / root cause: design

Terumo Cardiovascular Systems Corp began a recall of Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for... on . FDA classified it as Class II and gives the reason as "The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1828-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed it.
Product code
DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Reason for recall
The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.
Root cause tag
design
FDA root cause
Component design/selection
Action
Consignees were notified of the recall by phone beginning on 5/7/2015 and a Urgent Medical Device Recall letter, dated 5/7/2015. The letter identified the affected device, reason for recall, potential hazard, and correction. Customers were asked to review the notice and ensure that all users are aware of the issue. In addition, they are to confirm the quantity of procedure kits for return; and, complete and return the Customer Response Form. Customers can call Terumo CVS Customer Service at 1-800-521-2818 with questions.
Quantity in commerce
163 devices
Distribution
Distributed to the states of MO and FL.
Product codes and lots
Part Number: 66086-01: Lot Numbers: RG09, RM20 and Part Number: 70700-03: Lot Numbers: QL09, RH07, RL15, QG17, RD03

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1828-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1828-2015Class IITerminatedVoluntary: Firm initiated