Z-1942-2015Class II

EasyLink Data Management System; a clinical data management system that assists medical laboratory...

Siemens Healthcare Diagnostics Inc. / initiated 2015-04-29 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of EasyLink Data Management System; a clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions... on . FDA classified it as Class II and gives the reason as "System Limitations and Software Issues related to the following features: Custom reference ranges, User-initiated sample and result actions, Auto-verification and quality control rules, and Sample reports". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1942-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
EasyLink Data Management System; a clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
System Limitations and Software Issues related to the following features: Custom reference ranges, User-initiated sample and result actions, Auto-verification and quality control rules, and Sample reports.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent an Urgent Field Safety Notice (UMDC # ISW-15-01.A.US) on May 5, 2015 to all affected customers. The letter instructed customers to follow specific work around instructions, review this letter with their Medical Director, complete and return the Field Correction Effectiveness Check attached to this letter within 30 days, retain the letter with laboratory records, and forward it to those who may have received this product. For questions regarding this recall call 800-441-9250.
Quantity in commerce
1355
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Austria, Belgium, Canada, Switzerland, Germany, Denmark, Spain, France, Hong Kong, Italy, Japan, Korea, Malaysia, Netherlands, Norway, New Zealand, Philippines, Portugal, Saudi-Arabia, Slovenia, and Taiwan.
Product codes and lots
Catalog Numbers: 10444799, 10444800, 10487219

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1942-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1942-2015Class IITerminatedVoluntary: Firm initiated