Z-1682-2015Class II

Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical...

Philips North America LLC / initiated 2015-03-11 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical Systems. The device is a diagnostic imaging... on . FDA classified it as Class II and gives the reason as "It was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom. This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1682-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Ingenuity TF Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI), 882380, Philips Medical Systems. The device is a diagnostic imaging system for fixed installations that combines Positron Emission Tomography (PET) and Magnetic Resonance Imaging (MRI).
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
It was noticed that the spectroscopy voxel map did not align with the anatomical images of the phantom. This resulted in incorrect orientation of the Chemical Shift Imaging (CSI) data.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
On 3/11/2015 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected device, explained the reason for the recall, how to identify affected product, advice on actions to be taken by customer/user, and any actions planned by the firm. Customers were advised to use software provided by Philips to view spectroscopy images; and, not to export and view images using third party software. The firm stated that it will be issuing a software update. If further information or support concerning the issue is needed, customers are asked to contact their local Philips representative or local Philips Healthcare office.
Quantity in commerce
10 units
Distribution
Worldwide Distribution -- US, including the states of NY, OH, & TX; and, the countries of Finland, Germany, Japan, Netherlands, South Korea, Spain, & Switzerland.
Product codes and lots
Serial #'s: 11005, 11006, 11007, 11008, 11009, 11010, 11011, 11012, 11013, 11014 & 24061.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1682-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1682-2015Class IITerminatedVoluntary: Firm initiated