Z-1695-2015Class II

Brilliance iCT Computed Tomography X-ray system

Philips North America LLC / initiated 2015-04-13 / Terminated / root cause: supplier

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT Computed Tomography X-ray system on . FDA classified it as Class II and gives the reason as "A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1695-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance iCT Computed Tomography X-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
Root cause tag
supplier
FDA root cause
Other
Action
Philips sent an Urgent - Field Safety Notice dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Healthcare is recommending to stop the use of your system until the X-ray tube mounting fasteners are replaced. In addition, a Philips Field Service Engineer (FSE) will service your system via a Mandatory Field Change Order (FCO 72800638). A Philips FSE will contact you to schedule this appointment. If you need any further information or support concerning this issue, please contact your local Philips representative or your local Philips Healthcare office. This notice has been submitted to the appropriate Regulatory Agency. Philips apologizes for any inconveniences this problem has caused your organization. For further questions please call (440) 483-7600.
Quantity in commerce
2
Distribution
Internationally to the countries of China and Norway.
Product codes and lots
Model# 728306; Serial #'s: 85019 & 85023.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1695-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1695-2015Class IITerminatedVoluntary: Firm initiated