Z-1695-2015Class II
Brilliance iCT Computed Tomography X-ray system
Philips North America LLC / initiated 2015-04-13 / Terminated / root cause: supplier
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance iCT Computed Tomography X-ray system on . FDA classified it as Class II and gives the reason as "A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1695-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance iCT Computed Tomography X-ray system
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- A calibrated torque wrench malfunctioned at the supplier leading to possible incorrect torque of x-ray tube mounting fasteners.
- Root cause tag
- supplier
- FDA root cause
- Other
- Action
- Philips sent an Urgent - Field Safety Notice dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Healthcare is recommending to stop the use of your system until the X-ray tube mounting fasteners are replaced. In addition, a Philips Field Service Engineer (FSE) will service your system via a Mandatory Field Change Order (FCO 72800638). A Philips FSE will contact you to schedule this appointment. If you need any further information or support concerning this issue, please contact your local Philips representative or your local Philips Healthcare office. This notice has been submitted to the appropriate Regulatory Agency. Philips apologizes for any inconveniences this problem has caused your organization. For further questions please call (440) 483-7600.
- Quantity in commerce
- 2
- Distribution
- Internationally to the countries of China and Norway.
- Product codes and lots
- Model# 728306; Serial #'s: 85019 & 85023.