Z-1608-2015Class II

Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central...

GE Healthcare / initiated 2015-05-08 / Terminated / root cause: software

Spacelabs Healthcare Inc began a recall of Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs... on . FDA classified it as Class II and gives the reason as "The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1608-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Spacelabs qube Compact Monitor, Model 91390. The monitor is functioning as either bedside or central monitors; passively displays data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
The input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated in 2012 (Z-0403-2013).
Root cause tag
software
FDA root cause
Software design
Action
The firm sent the Urgent Medical Device Correction- Spacelabs Healthcare Inc.-qube Compact Monitor, Model 91390, Power Input PCBA Failure letter, dated 8 May 2015, via priority service mail to US consignees. This letter described the affected system, problem, actions to be taken by users and corrective actions conducted by Spacelabs. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
Quantity in commerce
5 units distributed in the US
Distribution
distributed in North Carolina, Texas, and Wyoming
Product codes and lots
Serial Numbers:  1390-000054,  1390-000064,  1390-000313,  1390-000047,  1390-000898.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1608-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1608-2015Class IITerminatedVoluntary: Firm initiated