Z-1734-2015Class II

AccuChek Spirit insulin infusion pump

Roche Diabetes Care, Inc. / initiated 2015-04-22 / Terminated / root cause: design

Roche Diabetes Care, Inc. began a recall of AccuChek Spirit insulin infusion pump on . FDA classified it as Class II and gives the reason as "Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1734-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diabetes Care, Inc.
Product
AccuChek Spirit insulin infusion pump.
Product code
LZG Pump, Infusion, Insulin
Reason for recall
Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and El0 error messages. If the user does not act upon the E6 and E10 error messages appropriately, insulin delivery will be stopped and if unnoticed may lead to severe hyperglycemia u
Root cause tag
design
FDA root cause
Device Design
Action
Roche sent an URGENT MEDICAL DEVICE CORRECTION notification dated April 22, 2015, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. The recall notification included a description of the reason for the recall, customer responsibilities, training chart, and instructions for responding to the formal recall notification. Please contact the ACCU-CHEK Customer Care Service Center, 24 hours a day, seven days a week at 1-800-688-4578, if you have questions about the information contained in this notification.
Quantity in commerce
35,795
Distribution
Nationwide Distribution
Product codes and lots
All Serial Numbers  Models: 1. 05458544001, Accu-Chek Combo Kit mg US/English version. 2. 05504686001, Accu-Chek Combo Kit mg US/Spanish version. 3. 05391512001, Accu-Chek Spirit Combo Replacement Pump US/EN. 4. 05391750001, Accu-Chek Spirit Combo Demo Pump US. 5. 04759702001, Accu-Chek Spirit Pump Kit US/CA Back Up. 6. 04540492001, Accu-Chek Spirit Pump Kit US/CA. 7. 04759729001, Accu-Chek Spirit Replacement Pump.