Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002...
Hamilton Medical, Inc. began a recall of Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and... on . FDA classified it as Class II and gives the reason as "Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2057-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hamilton Medical AG
- Product
- Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- *** Please note *** This is not a new recall. An error occurred, and this recall was originally classified as a class I on July 10, 2015. No further action is required. Hamilton Medical AG notified their customers with a letter dated February 19, 2015 and had a fix (SW 2.42) that was available as of February 20th 2015. Hamilton Medical proceeded to send out notification letters the 24th of March 2015 once they compiled the complete and validated list of customers that had the affected software. Customers with questions were instructed to call 800-426-6331, ext.215.
- Quantity in commerce
- 248
- Distribution
- Nationwide Distribution
- Product codes and lots
- Software version 2.20 and 2.41,
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2057-2015 | Class II | Terminated | Voluntary: Firm initiated |