Z-2057-2015Class II

Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002...

Hamilton Medical AG / initiated 2015-03-24 / Terminated / root cause: design

Hamilton Medical, Inc. began a recall of Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and... on . FDA classified it as Class II and gives the reason as "Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2057-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hamilton Medical AG
Product
Hamilton-G5 Ventilators with software versions V2.40/2.41 Catalog numbers for G5: 159001 and 159002. Software: 159700.. Intensive care ventilation of adult and pediatric patients and optionally infant and neonatal patients.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Customer reports that the ventilator display can freeze. Ventilation continues but the information is no longer displayed on the screen and the user can no longer operate the device.
Root cause tag
design
FDA root cause
Device Design
Action
*** Please note *** This is not a new recall. An error occurred, and this recall was originally classified as a class I on July 10, 2015. No further action is required. Hamilton Medical AG notified their customers with a letter dated February 19, 2015 and had a fix (SW 2.42) that was available as of February 20th 2015. Hamilton Medical proceeded to send out notification letters the 24th of March 2015 once they compiled the complete and validated list of customers that had the affected software. Customers with questions were instructed to call 800-426-6331, ext.215.
Quantity in commerce
248
Distribution
Nationwide Distribution
Product codes and lots
Software version 2.20 and 2.41,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2057-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2057-2015Class IITerminatedVoluntary: Firm initiated