Z-1726-2015Class I

HeartWare Ventricular Assist System (HVAD). Catalog #: 1101, 1103, 1100, 1101, 1102, 1104, 1104JP, and 1205...

HeartWare, Inc. / initiated 2015-04-29 / Open, Classified / root cause: design

HeartWare Inc began a recall of HeartWare Ventricular Assist System (HVAD). Catalog #: 1101, 1103, 1100, 1101, 1102, 1104, 1104JP, and 1205. Intended for use as a bridge to cardiac... on . FDA classified it as Class I and gives the reason as "Complaints with the HVAD Retraction of Pins within the driveline connector". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1726-2015
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
HeartWare, Inc.
Product
HeartWare Ventricular Assist System (HVAD). Catalog #: 1101, 1103, 1100, 1101, 1102, 1104, 1104JP, and 1205. Intended for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage heart failure.
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
Complaints with the HVAD Retraction of Pins within the driveline connector.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Medical Device Correction letters, dated APR2015, were sent to users. The letters identified the affected device, as well as the reason for the recall and risks involved. Users were asked to review and familiarize themselves with the notice and "Patient Communication." The notice should be forwarded to anyone within customers' organization that need to be aware. Customers are to identify their patients who are currently supported by the HVAD System and distribute the "Patient Communication" to them. A HeartWare representative can assist with that process is needed. Customers should continue to reinforce the messages in the notice with their patients. The "Acknowledgement and Completion Form" should be completed, signed, and returned within 30 days of receipt of the notice. Questions or concerns should be directed to a local HeartWare representative. Clinical Support is also available at 1-888-494-6365.
Quantity in commerce
3,747 currently in use
Distribution
Worldwide Distribution -- US, including the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, UT, and VA; and, the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Korea, Lebanon, Lithuania, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, and UK.
Product codes and lots
Serial #: ALL HeartWare HVAD systems currently in use.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1726-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1726-2015Class IOngoingVoluntary: Firm initiated