Z-1801-2015Class II

T.A.L.O.N (Tactically Advanced Lifesaving Intraosseous Needle) 15G x 38.5 mm 10 NEEDLE SET STERILE EO Single...

Teleflex Medical / initiated 2015-05-12 / Terminated / root cause: design

Vidacare Corporation began a recall of T.A.L.O.N (Tactically Advanced Lifesaving Intraosseous Needle) 15G x 38.5 mm 10 NEEDLE SET STERILE EO Single Use Rx only Vidacare Corporation on . FDA classified it as Class II and gives the reason as "Potential breach in sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1801-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
T.A.L.O.N (Tactically Advanced Lifesaving Intraosseous Needle) 15G x 38.5 mm 10 NEEDLE SET STERILE EO Single Use Rx only Vidacare Corporation
Product code
FMI Needle, Hypodermic, Single Lumen
Reason for recall
Potential breach in sterility
Root cause tag
design
FDA root cause
Device Design
Action
Teleflex sent an Urgent Medical Recall Notification dated May 12, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter directed the customers to quarantine the affected product and return it to the manufacturer. If you have any further questions please call 1-866-246-6990.
Quantity in commerce
2,104 boxes
Distribution
US Distribution to the states of : NC, MD, MS, TX and MO., and Internationally to Canada and Denmark.
Product codes and lots
3103376, 3103377, 3103378, 3524557, 3547157, 3693167, 3843307, 3868998, 3986753, 3991994, 3991995, 3999583, 4011996, 4023525, 4040723, 4040724

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1801-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1801-2015Class IITerminatedVoluntary: Firm initiated