Z-1801-2015Class II
T.A.L.O.N (Tactically Advanced Lifesaving Intraosseous Needle) 15G x 38.5 mm 10 NEEDLE SET STERILE EO Single...
Teleflex Medical / initiated 2015-05-12 / Terminated / root cause: design
Vidacare Corporation began a recall of T.A.L.O.N (Tactically Advanced Lifesaving Intraosseous Needle) 15G x 38.5 mm 10 NEEDLE SET STERILE EO Single Use Rx only Vidacare Corporation on . FDA classified it as Class II and gives the reason as "Potential breach in sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1801-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- T.A.L.O.N (Tactically Advanced Lifesaving Intraosseous Needle) 15G x 38.5 mm 10 NEEDLE SET STERILE EO Single Use Rx only Vidacare Corporation
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- Potential breach in sterility
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Teleflex sent an Urgent Medical Recall Notification dated May 12, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter directed the customers to quarantine the affected product and return it to the manufacturer. If you have any further questions please call 1-866-246-6990.
- Quantity in commerce
- 2,104 boxes
- Distribution
- US Distribution to the states of : NC, MD, MS, TX and MO., and Internationally to Canada and Denmark.
- Product codes and lots
- 3103376, 3103377, 3103378, 3524557, 3547157, 3693167, 3843307, 3868998, 3986753, 3991994, 3991995, 3999583, 4011996, 4023525, 4040723, 4040724
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1801-2015 | Class II | Terminated | Voluntary: Firm initiated |