Unilateral external fixation ankle clamp
Biomet, Inc. began a recall of Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of... on . FDA classified it as Class II and gives the reason as "Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1741-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Unilateral external fixation ankle clamp Product Usage: Unilateral external fixation for use treatment of bone conditions amenable to treatment by use of external fixation modality
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- Ankle clamp was assembled incorrectly; the offset of the pin to the pivot point is reversed. This prevents complete range of motion and if pushed too far may loosen the pin from the bone. A revision surgery may be necessary to adequately reconnect the fixator to the bone.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Biomet sent an URGENT MEDICAL DEVICE RECALL NOTICE letter dated May 1, 2015 to the affected distributors and implanting surgeons. The recall letter included a description of the reason for the recall, affected product, consignee responsibilities, and distributor instructions for responding to the formal recall notification. Consignees were instructed to return the affected product and continued monitoring of patients. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
- Quantity in commerce
- 19
- Distribution
- US Nationwide Distribution in the states of CA, TX, FL, NC, MI, GA.
- Product codes and lots
- Part 14-450510 Lots:268000, 466170, and 813950
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1741-2015 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY