510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1819-2015Class II

ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with...

Orthofix SRL / initiated 2015-05-11 / Terminated / root cause: design

Orthofix, Inc began a recall of ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure... on . FDA classified it as Class II and gives the reason as "There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1819-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.
Product code
FZX Guide, Surgical, Instrument
Reason for recall
There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.
Root cause tag
design
FDA root cause
Device Design
Action
Orthofix sent an Urgent Medical Device Product Advisory Notification letter dated May 11, 2015, to the affected distributors and hospitals. The consignees were directed to inspect thier inventory and perform checks before surgical use of the drill guides. Customers were instructed to provide a copy of the Product Advisory Notification letter if product was further distributed. Customers with questions were instructed to contact their local Orthofix representative and to for returns they should contact Orthofix customer service representative at 888-298-5700.
Quantity in commerce
36 units
Distribution
Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.
Product codes and lots
All lot codes

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1819-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1819-2015Class IITerminatedVoluntary: Firm initiated