Z-1810-2015Class II

Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera...

Stryker Corp. / initiated 2015-05-04 / Terminated / root cause: manufacturing process

Stryker Instruments Div. of Stryker Corporation began a recall of Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation... on . FDA classified it as Class II and gives the reason as "Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1810-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Navigation System II-Cart, 7700-100-000 with Articulated Arm Camera, 7700-103-001 and Articulated Arm Camera, 7700-103-001U Product Usage: The Navigation System II-Cart is intended to be used as a platform for computer assisted surgery as part of the Stryker Navigation System. The intended use of the complete Stryker Navigation System is defined by the Stryker application modules.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Stryker initiated a device recall of the Navigation System II-Cart, Camera Articulated Arm because there is a potential that the weld seam between the main stud and the two flaps on the Articulated Arm was not welded in the correct location, which may result in a potential failure of the arm joint.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker sent an Urgent Medical Device Recall Notification letter dated May 18, 2013 to Domestic customers notice via FedEx overnight and Sales Representatives via email. The International Subsidiaries will also be notified via email. Customers are instructed to immediately review the Recall Notification adn complete the following: Immediately check all stock areas and/or operating room storage for affected equipment found. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many affected items are currently in inventory. Please complete and return the BRF even if they dont have any affected product on hand. If customers have further distributed this product, they are asked to forward this letter and the attached BRF to all affected locations. Please indicate each location and serial number(s) on the BRF. Fax (866-521-2762) or email ([email protected]) the completed BRF to Stryker Instruments Regulatory Department, Attn: Kelly Jo Davis. A Stryker Representative will contact the facility to set up a time to perform the additional inspection of the weld seam on the Navigation System II-Cart, Articulated Arm Camera.
Quantity in commerce
72
Distribution
Worldwide Distribution - US Nationwide in the states of AR, CO, FL, GA, IN, KS, KY, MA, MI, NC, NE, NJ, NY, OH, OK, PA, TX, WA, WY and the countries of Canada, Europe, and Asia.
Product codes and lots
Serial numbers 100019, 100984, 100965, 100295, 100979, 100773, 100946, 100797, 100947, 100956, 100966, 100649, 100967, 100944, 100941, 100945, 100970, 100950, 100976, 100951, 100728, 100952, 100248, 100953, 100641, 100954, 100463, 100955, 100981, 100957, 100978, 100958, 100106, 100959, 100939, 100960, 100973, 100961, 100236, 100962, 100237, 100963, 100417, 100964. 100940, 100968, 100975, 100969, 100977, 100972, 100105, 100974, 100214, 100980, 100492, 100982, 100493, 100983, 100948, 100986, 100704, 100987, 100949, 100988, 100227, 100989, 100971, 100133, 100134, 100942.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1810-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1810-2015Class IITerminatedVoluntary: Firm initiated