Z-2062-2015Class II
0210-312-000-Woundcare Replacement kit with soft tip; 0210-318-000; Woundcare Replacement with fan tip...
Stryker Corp. / initiated 2015-05-08 / Terminated / root cause: packaging
Stryker Instruments Div. of Stryker Corporation began a recall of 0210-312-000-Woundcare Replacement kit with soft tip; 0210-318-000; Woundcare Replacement with fan tip; 0210-318-100- Woundcare Replacement Kit... on . FDA classified it as Class II and gives the reason as "Potential sterility breach in the packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2062-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- 0210-312-000-Woundcare Replacement kit with soft tip; 0210-318-000; Woundcare Replacement with fan tip; 0210-318-100- Woundcare Replacement Kit; 0210-318-200-Woundcare Replacement with retract coax tip. Accessory to Stryker InterPulse Irrigation System, used for wound debridement, soft tissue debridement, and cleansing of medical, clinical, or surgical sites.
- Product code
- FQH Lavage, Jet
- Reason for recall
- Potential sterility breach in the packaging.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Stryker sent Customer Notification Letters on 05/20/2015 through Certified mail via USPS. Customers are asked to do the following: Immediately review this Recall Notification. Immediately check all stock areas and/or operating room storage and quarantine any affected product found. Complete the enclosed Business Reply Form (BRF) to confirm receipt of this notification and identify how many, if any, affected items are currently in your inventory. Please complete and return the BRF even if you do not have any affected product on hand. If you have further distributed this product, please forward this letter and the attached Business Reply Form (BRF) to all affected locations. Please indicate each location on the BRF. Fax (866-521-2762) or email ([email protected]) the completed Business Reply Form to Stryker Instruments Regulatory Department. Upon receipt of the Business Reply Form, if you have recalled product on hand, Stryker will email a pre-paid shipper. This shipper can be used to return the recalled WoundCare Replacement kits to Stryker.
- Quantity in commerce
- 5,328 each (444 boxes)
- Distribution
- Nationwide Distribution-including the states of FL, IA, IL, IN, LA, MD, MN, MO, NC, NM, NY, PA, TN, TX, VA, and WI.
- Product codes and lots
- 0210-312-000: 12138012, 12203012, 12229012, 12256012, 12293012, 12314012, 12338012, 13086012, 13140012, 13192012; 0210-318-000: 14140012,14258012; 0210-318-100: 12294012,13051012; 0210-318-200: 12143012, 12293012, 13066012, 14037012, 14112012
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2062-2015 | Class II | Terminated | Voluntary: Firm initiated |