Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540...
Medtronic Navigation, Inc. began a recall of Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product... on . FDA classified it as Class II and gives the reason as "Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1794-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product is either packaged with other instruments or into a holding tray. The Passive Planar Blunt Probe is an optically tracked, reusable instrument used in navigated surgery and is included with Medtronic cranial instrument kits. It is designed for use in patient registration and palpating brain tissue during surgical procedures.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip.
- Root cause tag
- manufacturing process
- FDA root cause
- Mixed-up of materials/components
- Action
- Medtronic sent an Medical Device Safety Alert letter dated May 5, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken by the user: "Examine your inventory and if any of the affected products are found, please quarantine them for return to Medtronic. "A Medtronic representative will follow-up to arrange for removal and replacement. If you have any questions regarding this action, please contact Medtronic Navigation Technical Services at 1-800-595-9709.
- Quantity in commerce
- 18
- Distribution
- US Distribution to the states of : CA, MO, LA and PA., and Internationally to Ireland.
- Product codes and lots
- lot number 150223
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1794-2015 | Class II | Terminated | Voluntary: Firm initiated |