Z-1794-2015Class II

Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540...

Medtronic Navigation, Inc / initiated 2015-05-05 / Terminated / root cause: manufacturing process

Medtronic Navigation, Inc. began a recall of Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product... on . FDA classified it as Class II and gives the reason as "Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1794-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Passive Planar Blunt Probe Part Number 960-556, also included in Part Number Kits: 9732540, 9732622, 9733174, 9733935. As part of kits, the product is either packaged with other instruments or into a holding tray. The Passive Planar Blunt Probe is an optically tracked, reusable instrument used in navigated surgery and is included with Medtronic cranial instrument kits. It is designed for use in patient registration and palpating brain tissue during surgical procedures.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Passive Planar Blunt Probe (Lot# 150223) was manufactured with a sharp tip rather than a blunt tip.
Root cause tag
manufacturing process
FDA root cause
Mixed-up of materials/components
Action
Medtronic sent an Medical Device Safety Alert letter dated May 5, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken by the user: "Examine your inventory and if any of the affected products are found, please quarantine them for return to Medtronic. "A Medtronic representative will follow-up to arrange for removal and replacement. If you have any questions regarding this action, please contact Medtronic Navigation Technical Services at 1-800-595-9709.
Quantity in commerce
18
Distribution
US Distribution to the states of : CA, MO, LA and PA., and Internationally to Ireland.
Product codes and lots
lot number 150223

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1794-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1794-2015Class IITerminatedVoluntary: Firm initiated