Z-1679-2015Class II

RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment

RaySearch Laboratories AB (publ) / initiated 2015-05-08 / Terminated / root cause: software

RAYSEARCH LABORATORIES AB began a recall of RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and... on . FDA classified it as Class II and gives the reason as "RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1679-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
RaySearch RayStation 2.5, 3.0, 3.5 and 4.0; Radiation Therapy Treatment Product Usage: RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
RaySearch Laboratories AB became aware of an issue with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing.
Root cause tag
software
FDA root cause
Software Design Change
Action
RaySearch sent a Device Correction letter dated May 8, 2015 via e-mailed to their customers. The letter identified the affected product, problem, actions to be taken by users and contact information.
Quantity in commerce
Domestic: 539 units
Distribution
US Nationwide Distribution
Product codes and lots
Software Version (Build numbers) 2.5.1.89, 3.0.0.251, 3.5.0.16, 3.5.1.6, 4.0.0.14, 4.0.1.4, 4.0.2.9, or 4.0.3.4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1679-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1679-2015Class IITerminatedVoluntary: Firm initiated