Z-1638-2015Class II
GE OEC Brivo 865 plus, Mobile X-ray system laser aimer assemblies (part number 520002DW)
GE Healthcare / initiated 2014-05-22 / Terminated
GE OEC Medical Systems, Inc began a recall of GE OEC Brivo 865 plus, Mobile X-ray system laser aimer assemblies (part number 520002DW) on . FDA classified it as Class II and gives the reason as "Laser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1638-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE OEC Brivo 865 plus, Mobile X-ray system laser aimer assemblies (part number 520002DW)
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- Laser aimer assemblies were shipped without certification of the overall assembly or submitting a laser product report and therefore do not comply with the US regulations for laser products.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Healthcare Surgery Planned Action(s) to repair the defect or to bring the product into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 9 units consisted of the following: 1. Replacing the tube side cover with the laser aimer assembly with a tube side cover without a laser aimer assembly. This correction brought the systems into compliance with applicable US regulations. For further questions please call (801) 536-4615.
- Quantity in commerce
- 9
- Distribution
- US Distribution to the states of: MD, NC, MT, NJ, UT and NY
- Product codes and lots
- OEC Brivo 865 Plus serial numbers B4S13061, B4S14006, B4S14007, B4S13033, B4S13032, B4S13038, B4S13039, B4S13024, AND B4S13023.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1638-2015 | Class II | Terminated | Voluntary: Firm initiated |