ICT Serum Calibrator REF 146-03. For use in the calibration of the serum Sodium, Potassium, and Chloride...
Abbott Laboratories began a recall of ICT Serum Calibrator REF 146-03. For use in the calibration of the serum Sodium, Potassium, and Chloride assays on . FDA classified it as Class II and gives the reason as "ICT Serum Calibrator may generate lower than expected Potassium Quality Control (QC) and patient results on the Abbott Architect cSystem". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1932-2015
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ICT Serum Calibrator REF 146-03. For use in the calibration of the serum Sodium, Potassium, and Chloride assays.
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- ICT Serum Calibrator may generate lower than expected Potassium Quality Control (QC) and patient results on the Abbott Architect cSystem.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customers who received shipments of the affected lots were sent a Product Correction letter on July 22, 2014.
- Quantity in commerce
- 11,215 units
- Distribution
- Worldwide Distribution-US (nationwide) including Puerto Rico, and the countries of Afghanistan, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Bermuda, Bosnia & Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Curacao, North Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malawi, Malaysia, Mali, Mauritania, Mauritius, Mexico, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, Turks & Calicos, UAE, Uganda, United Kingdom, Uruguay, Venezuela, Vietnam, and Zimbabwe.
- Product codes and lots
- Lot numbers: 21399UN13 (Expiration date: 5AUG2014), 67873UN13 (Expiration date: 12NOV2014), 74710UN13 (Expiration date: 8DEC2014), 82008UN13 (Expiration date: 14JAN2015), 88884UN13 (Expiration date: 4FEB2015)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1932-2015 | Class II | Terminated | Voluntary: Firm initiated |