Z-1688-2015Class II

Base for Terumo Advanced Perfusion System 1, 220/240V The Terumo Advanced Perfusion System 1 is indicated for...

Terumo Cardiovascular Systems Corporation / initiated 2015-05-01 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Base for Terumo Advanced Perfusion System 1, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal... on . FDA classified it as Class II and gives the reason as "Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1688-2015
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Base for Terumo Advanced Perfusion System 1, 220/240V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to preven
Root cause tag
design
FDA root cause
Device Design
Action
On 5/1/2015 Terumo issued an URGENT MEDICAL DEVICE CORRECTION letter to users explaining that Terumo has identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to prevent inadvertent shut off. - Replacing the On/Off power switch with a new design to prevent switch failure. A Terumo Field Service Representative will contact users to schedule the field correction activities once the activities are ready to be implemented. CUSTOMER INSTRUCTIONS: Review this Medical Device Correction and assure that all users have received notice of this issue. - Complete and return the enclosed Customer Response Form. - Terumo CVS will contact users to schedule the field correction activities. We encourage you to contact us with any questions or concerns: Terumo CVS Customer Service 1-800-521-2818 Fax 1-734-741-6149 Customer Service Hours: Monday Friday, 8 a.m. 6 p.m. ET
Quantity in commerce
1713 units total
Distribution
Worldwide Distribution-US (nationwide) including DC and the states of AL , AR , AZ , CA , CO , CT , DE , FL , GA , HI , IA , ID , IL , IN , KY , LA , MA , MD , ME , MI , MN , MO , MS , MT , NC , ND , NE , NJ , NM , NV , NY , OH , OK , OR , PA , SC , TN , TX , UT , VA , VT , WA , WI , and WV, and the countries of Mexico , AUSTRALIA , UNITED ARAB EMIRATES (UAE) , Hong Kong , Indonesia , Singapore , Taiwan , Thailand , Philippines , South Korea , CHILE , Vietnam , Malaysia , BELGIUM , Japan , and CANADA.
Product codes and lots
Catalog number: 801764 and All serial number units

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1688-2015
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1687-2015Class IITerminatedVoluntary: Firm initiated
Z-1688-2015Class IITerminatedVoluntary: Firm initiated